Technical parameters of EH-737 medical intelligent infusion pump | |
---|---|
pumping method | Half Squeeze Finger Peristaltic Pump |
Applicable infusion set | Special or high-quality PVC and TPE infusion sets with an outer diameter of 3.4mm to 4.5mm. |
Four infusion modes | Flow rate - total formula time-total formula Drip rate - total amount Weight Mode |
Flow rate - total | Flow rate: 1.0~1200.0 ml/h, 0.1ml/h increments |
time-total | Time: 1 second to 99:59:59 in increments of 1 second |
Drip rate - total amount | Drip rate: 1~333 drops/min, 1 drop increments |
Preset/Total | Preset volume: 0.1~9999.9ml, 0.1ml increment Cumulative volume: 0~9999.9ml, can be reset |
Weight Mode | Dosage: 0.1 ~ 999.9 mg, 0.1 mg incremental Mg Dosage: 0.01 to 999.99 mg/kg/h, in increments of 0.01 mg/kg/h Weight: 1 to 300 kg, in 1 kg increments |
KVO function | KVO flow rate: 1.0 to 5.0 ml/h, in increments of 0.1 ml/h |
Bolus function | Bolus flow rate: 50.0 to 800.0 ml/h, in increments of 0.1 ml/h Bolus capacity: 0.1~99.9 ml/h, 0.1ml increments |
Infusion accuracy | Within 1.0~1200.0 ml/h, the measured accuracy is better than ±5% (use calibrated special or high-quality PVC, TPE infusion set) |
Call the police | 1. Bubble alarm, the sensitivity is adjustable in 3 gears; 2. Bubble sensor damage alarm; 3.1. The upper end pipeline blockage alarm, -20~-60 kPa, the sensitivity is adjustable in 5 gears; 3.2. The initial pressure of the upper pipeline is abnormal (the upper pressure sensor is damaged) alarm; 4.1. The lower end pipeline blockage alarm, 20~140 kPa, the sensitivity is adjustable in 5 gears; 4.2. The initial pressure of the lower pipeline is abnormal (the lower pressure sensor is damaged) alarm; 5.1. The abnormal rotation of the peristaltic pump alarms; 5.2. Peristaltic pump reverse (infusion pumping) alarm; 6. Abnormal door opening alarm; 7. Insufficient battery power and battery failure alarm; 8. Pause timeout alarm; 9. Alarm when infusion is completed; 10. Mains abnormality prompt; 11. The microprocessor alarms abnormally and shuts down. |
Data function | 1: Calibration memory function: automatically record the pipeline calibration data, which is convenient for subsequent use; 2: Parameter memory function: automatically record the last infusion parameters, which is convenient for subsequent use; 3: History record (SOE function): It can record the last 1000 infusion events (infusion start/stop time, flow rate, stop reason, etc.). |
power supply | AC power supply: 100~240V, 50/60Hz Built-in rechargeable NiMH battery Power consumption: ≤20VA |
Appearance and Weight | Width 130mm×height 206mm×thickness 118mm The weight of the whole machine is less than 1.5kg |